Operating Instructions Hanging Balance
1 Technical
Data
nature and scope of this check. Information is available on
KERN
CH15K20 CH50K50 CH50K100
KERN's home page (www.kern-sohn.com) with regard to the
monitoring of balance test substances and the test weights
required for this. Test weights and balances can be adjusted
Readout
20 g
50g
100g
quickly and at a reasonable price in KERN's accredited DKD
Weighing range
15 kg
50kg
50kg
calibration laboratory (return to national normal).
2
Fundamental information (general)
3
Fundamental safety information
2.1 Intended
use
The balance you have acquired serves to determine the
Do not use the hanging balance to transport loads.
weighing value of the material to be weighed. It is intended to
be used as a "non-automatic" balance, i.e. the material to be
weighed is manually and carefully attached at the balance.
Prevent jolts, torsion and oscillation (e.g. by appending
The weighing value can be read off after a stable weighing
slopingly) of all kinds.
value has been obtained.
Never use the hanging balance over the maximum
2.2 Inappropriate
use
permitted weight (!!Danger of breaking!!).
Do not use the balance for dynamic weighing. In the event that
small quantities are removed or added to the material to be
Always make sure that there are
no living beings or
weighed, incorrect weighing results can be displayed due to
materials
below the load that could be injured or dam-
the "stability compensation" in the balance. (Example: Slow
aged.
draining off of liquid from a container suspended from the
balance)
Do not attach a continuous load. This can damage the meas-
The hanging electronic balances from the
KERN CH
series are only suitable for hand-held use (see photo).
uring unit as well as the parts, relevant to safety.
They are
not
suitable for hanging from a mechanical
Prevent jolts, torsion and oscillation (e.g. by appending slop-
hook, e.g. a crane hook.
ingly) of all kinds. Be sure to prevent overloading the balance
in excess of the stated maximum load (max.), minus any tare
weight that may possibly exist. This could damage the balance
(risk of breakage).
Important:
·
Always make sure that there are no people or materi-
als below the load that could be injured or damaged!
·
The balance is not suitable for weighing people. Do
not use as baby scales!
·
The balance does not comply with the medical prod-
uct law (MPG).
Never operate the balance in hazardous locations. The series
design is not explosion-proof. Structural alterations may not be
made to the balance. This can lead to incorrect weighing re-
sults, faults concerning safety regulations as well as to de-
struction of the balance. The balance may only be used in
compliance with the described guidelines. Varying areas of
application/planned use must be approved by KERN in writing.
2.3 Guarantee
The guarantee is not valid following
·
non-observation of our guidelines in the operating
instructions
·
use outside the described applications
·
alteration to or opening of the device
·
mechanical damage and damage caused by media,
3.1 Observe the information in the operating instructions
liquids
·
natural wear and tear
Please read the operating instructions carefully before erecting
·
inappropriate erection or electric installation
and commissioning, even if you already have experience with
·
overloading of the measuring equipment
KERN balances.
2.4
Monitoring the test substances
3.2 Staff
training
The metrology features of the balance and any possible avail-
The device may only be operated and looked after by trained
able adjusting weight must be checked at regular intervals
members of staff.
within the scope of quality assurance. For this purpose, the
answerable user must define a suitable interval as well as the
CH-BA-defsi-0422
5
6 Adjustment
4
Transport and storage
As the acceleration value due to gravity is not the same at
4.1 Acceptance
check
every location on earth, each balance must be coordinated
Please check the packaging immediately upon delivery and
in compliance with the underlying physical weighing principle -
the device during unpacking for any visible signs of external
to the existing acceleration due to gravity at its place of loca-
damage.
tion ( only if the balance has not already been adjusted to the
location in the factory). This adjustment process must be
carried out during the initial start-up, after change in location
4.2 Packaging
and variation of surrounding temperature. It is also recom-
Please retain all parts of the original packaging in case it
mendable to adjust the balance periodically during weighing
should be necessary to return items at any time.
operation in order to obtain exact measured values.
Only the original packaging should be used for return con-
signments. Before despatch, disconnect all attached cables
6.1 Adjusting
and loose/movable parts. Apply any intended transport secu-
rity devices. Secure all parts, e.g. glass windshield, weighing
Using a precision weight, the accuracy of the balance can be
plate, power unit etc., to prevent slipping and damage.
checked at any time and adjusted.
Adjustment procedure:
5
Unpacking, installation and commissioning
Check that the surrounding conditions are stable. A short
warm-up time of about 1 minutes are recommended for stabi-
5.1
Place of installation, place of use
lisation.
The balance is constructed in such a way that reliable weigh-
Press "
OFF
" key approx. 3 seconds, "
CAL
" and then the exact
ing results can be achieved under normal application condi-
weight of the adjusting weight will appear at the display.
tions.
Hang on the adjusting weight on the hook of the hanging
By selecting the correct application location
f
or your balance,
scale. "
F
" will appear a short time later, and then turn off.
you will be able to work quickly and precisely.
In case of an adjusting error or a wrong adjusting weight "
E
"
Therefore please observe the following at the
application
appears in the display.
location
:
· Avoid extreme heat as well as temperature fluctuation
KERN
CH15K20 CH50K50 CH50K100
caused by installing next to a radiator or in the direct
Adjusting weight
10kg (M3) 20kg (M3) 20kg (M3)
sunlight;
(not included))
· Protect the balance against direct draughts due to open
windows and doors;
· Avoid jarring during weighing;
·
7 Operation
Instruction
Protect the balance against high humidity, vapours and
dust;
· Do not expose the device to extreme dampness for longer
7.1 Weighing
periods of time. Inadmissible bedewing (condensation of
air moisture on the device) can occur if a cold device is
· The balance is switched on by pressing the "
ON/TARE
" key.
taken into a significantly warmer environment. In this case,
· The balance display shows
"0.00".
please acclimatise the device for approx. 2 hours at room
temperature after it has been disconnected from the bat-
· Hang on the weighing object carefully.
terys.
· Avoid static charging of the material to be weighed, weigh-
· If the weighing object is heavier than the weighing range the
ing container and windshield.
display indicates "
E"
(=Overload). Be sure to prevent over-
loading (chap. 2.2 "Inappropriate use").
Major display deviations (incorrect weighing results) are possi-
· The balance is switched off by pressing the "
OFF"
key once.
ble if electromagnetic fields occur as well as due to static
charging and instable power supply. It is then necessary to
change the location.
7.2 Taring
· Switch the balance on with "
ON"
key and wait for "0" to ap-
5.2 Unpacking
pear in the display. Hang on the tare load carefully and press
Carefully remove the balance from its packaging, remove the
the "
ON/TARE
" key again.
plastic wrapping and position the balance in its intended work-
ing location.
· The balance display becomes "
0.00".
The weight of the tare
load has however been memorised internally. The item to be
5.2.1
List of items supplied
weighed can now be placed carefully in the tare container
and the weight can be read off.
Standard accessories:
· Hanging balance
· If the "
ON/TARE
" key is pressed, the total tare value will
· Battery
appear. When the "
ON/TARE
" key is pressed again, "
0.00
"
· Operating instructions
appears on the display. The total tare will be memorised.
5.3
Battery operation and change
7.3 Data
HOLD
Push off the battery cover on the back of the balance in the
· When the "
HOLD
" key is pressed, the display value (even
direction of the arrow. Insert 9 V battery. Replace the battery
when the weight values vary) shown on the display will be
cover. When the batteries are exhausted the message "LO"
held 5 seconds.
appears in the balance display. Press the OFF switch and
replace the batteries immediately. To preserve battery life the
balance switches off automatically after 4 minutes if no weigh-
ing has taken place.
6
CH-BA-defsi-0422
8
Maintenance, upkeep, disposal
8.1 Cleaning
Only use a cloth dampened with mild suds and not aggressive
cleaning agents (solvents or similar). Please ensure that fluids
are not able to get into the device and rub off using a clean,
soft cloth. Loose sample residue/powder can be removed
carefully using a brush or hand vacuum cleaner.
8.2 Maintenance,
upkeep
The operating company shall ensure that load-bearing com-
ponents are checked by a competent person at maximum
intervals of one year.
These checks can be necessary at intervals shorter than one
year depending on the conditions of use. This applies, for
example, in the case of particularly frequent use, increased
wear and corrosion.
The device may only be opened by trained service engineers
authorised by KERN.
Disconnect from the mains supply before opening.
8.3 Disposal
The operating company shall dispose of the packaging and
the device in compliance with the valid national or regional law
of the operating location.
9 Troubleshooting
The balance should be switched off for a short time following
an interruption in the programme sequence. It is then neces-
sary to repeat the weighing process from the beginning.
Interruption Possible
cause
Weight display is
·
The balance is not switched on.
not illuminated.
·
The batteries are wrongly inserted, the
batteries are empty
·
No batteries are attached
The weight dis-
·
Draught/air movement
play changes
·
Table/floor vibrations
continually
· Vibrations on the balance suspension
·
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
The weighing
·
The balance display is not set to zero
result is obviously
incorrect
·
Adjustment is no longer correct.
·
Great fluctuations in temperature.
·
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Switch the balance off if other error messages should appear
and then switch on again. Contact the manufacturer if the error
message does not disappear.
CH-BA-defsi-0422
7